2.7.6 Measurement Accuracy Test
Measurement Conditions
• Measured CH is arbitrary (CH2 recommended). Also, please remember that the
measurement accuracy test is carried out at the two measurement intervals of 100 ms
and 200 ms, and the ranges and criteria differ for each.
• Set the dip switches on the -B12 and -B35 modules for 4-gauge.
• The duration of the measured data at the 100 ms interval should be five seconds or
more, and 10 seconds or more at 200 ms, and all measured data must be checked
against the test criteria.
• If ever a criterion is not met, the unit should be calibrated and retested.
• Before performing the measurement accuracy test, restore the initial balance values
(all measuring ranges) of the measurement channels from a PC via communication.
For instructions on restoring initial balance values through PC communication, see
the MX100 standard software user’s manual (IM MX180-01E) or the MW100 data
acquisition unit user’s manual (IM MW100-01E).
Connection
Wiring diagram for the -B12 and -B35 modules
• Because the gain error will be equal to the wiring resistance of the P and N lines,
wiring must be performed so that each wiring resistance is kept to 10 mΩ or less.
• In the figure above, two P and N lines each are connected to the Cal Box in order to
reduce wiring resistance. You do not necessarily have to connect two lines each.
2.7 Tests of the 4-CH Strain Input Module