Yokogawa PC-Based MX100 Uživatelský manuál Strana 24

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 141
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 23
2-10
IM MX180-01E
3.
Click the Current Configuration tab followed by the Reconfigure based on actual
config. button.
A dialog box with the message “Module settings which do not match the current software
configuration will be initialized.” appears.
Click here
Click here
Note
If the × mark remains even after reconfiguring the system numerous times, see the corrective
action described for “The connected input/output module is not detected” in section 5.1.
4.
Click the OK button.
The current module configuration is matched to the actual configuration.
Current module configuration
Setting the A/D Integration Time
1.
Click System to show the System setup screen.
2.
From the A/D Integration Time list box, select AUTO, 50 Hz, or 60 Hz.
2.2 Configuring the System, Setting the Write Mode of the CF Card, and Formatting the CF Card