11. PARAMETER LIST
List of Parameters for HART Communicator
PV –32400 to 32400 R Displays instantaneous flow rate in engineering unit. A20
Totl 0 to 999999 0 W Displays and presets forward direction totalized flow rate. A30
PV AO 2.40 to 21.6 mA R Displays current output. —
Span 0.0001 to 30000 1.0000 W Sets flow span in selected unit. b03
Status group 1 Error R Displays self-diagnostics result. —
µP fault “ON” shows error status.
EEPROM fault “OFF” shows normal status.
A/D fault See “8.2 Self-diagnostics Functions.”
Signal overflow
Coil open
Vel. span> 10 m/s
Vel. span< 0.3 m/s
Status group 2 P. span> 1000p/s R Ditto —
P. span> 500p/s
P. span> 25p/s
P. span> 15p/s
P. span> 10p/s
P. span> 5p/s
P. span< .0001p/s
T. span> 1000p/s
Status group 3 T. span< .0001p/s R Ditto —
Analog lmt error
Empty pipe
Multi range err
Coil short
EEPROM no set
EEPROM default
Dev id not entered
Self test — — W Excecutes self-diagnostics. —
Auto zero — — W Executes automatic zero adjustment. C02
Zero tuning (00) Inhibit Enable W Inhibits or enables automatic zero adjustment. C01
(01) Enable
D/A trim ZERO: 3.2 to 5.6 mA W Performs fine adjustments of zero and span of current output. —
SPAN: 18.4 to 21.6 mA
Scaled D/A trim ZERO: 3.2 to 5.6 mA W Performs fine adjustments of zero and span of scaled analog —
SPAN: 18.4 to 21.6 mA output.
Loop test 4 mA 4 mA W Sets test output value. H02
20 mA
Other (2.72 to 21.28 mA)
End
Status out (00) Normal Normal W Selects test status output. H03
(01) Closed (on)
(02) Open (off)
PV Unit gal/min m/s W Selects engineering unit of “PV”. Displays “Spcl” in case of —
L/min setting unit except 5 units on the left.
Cum/h
ft/s
m/s
(Spcl)