IM 1E10C1-E
8-4
8. ACTUAL OPERATON
8.2 Self-diagnostics Functions
The self-diagnostics function displays instrument internal errors, input/output signal
abnormalities, setting errors, and other problems.
When an alarm occurs, the alarm number that the error has occurred and normal data
are alternately indicated on the indicator of ADMAG SE.
However, alarms are only displayed during flow rate data display mode and when
parameter numbers are being changed in the setting mode. (Alarms are not displayed
when data items are being changed.)
The current output during the alarm occurrence can be selected from among the
followings. See “8.2.1 Output Status during Alarm Occurence.”
2.4mA or less
4mA
HOLD
21.6mA or more
When the BT200 is used, alarms are displayed in “A60 to N60:SELF CHECK
parameter.
When the HART Communicator is used, alarms are displayed in “Status group 1 to
3” parameters. There are two methods for self-diagnostics, every transmission and
manually executing the “Self test” command. See “8.2.2 Self-diagnostics Using
HART Communicator.”
When an error message appears, follow “8.2.3 Error Description and Countermea-
sure.”.
8.2.1 Output Status during Alarm Occurrence
--- GOOD Self test OK ON Normal Normal
--- ERROR Error OFF Fixed* Stop
Err-02 uP FAULT uP fault OFF 21.6mA or more** Stop
Err-03 EEPROM FAULT EEPROM fault OFF 21.6mA or more** Stop
Err-05 A/D FAULT A/D fault OFF Fixed* Stop
Err-06 SIGNAL OVERFLOW Signal overflow OFF Fixed* Stop
Err-07 COIL OPEN Coil open OFF Fixed* Stop
Err-08 SPAN VEL.>10m/s Vel. span>10m/s OFF Fixed* Stop
Err-09 SPAN VEL.<0.3m/s Vel. span<0.3m/s OFF Fixed* Stop
Err-10 P.SPAN>1000p/s P.span>1000p/s OFF Fixed* Stop
Err-11 P.SPAN>500p/s P.span>500p/s OFF Fixed* Stop
Err-12 P.SPAN>25p/s P.span>25p/s OFF Fixed* Stop
Err-13 P.SPAN>15p/s P.span>15p/s OFF Fixed* Stop
Err-14 P.SPAN>10p/s P.span>10p/s OFF Fixed* Stop
Err-15 P.SPAN>5p/s P.span>5p/s OFF Fixed* Stop
Err-16 P.SPAN<.0001p/s P.span<.0001p/s OFF Fixed* Stop
Err-17 T.SPAN>1000p/s T.span>1000p/s OFF Fixed* Stop
Err-18 T.SPAN>.0001p/s T.span<.0001p/s OFF Fixed* Stop
Err-19 4-20 LMT ERROR Analog lmt error OFF Fixed* Stop
Err-20 EMPTY PIPE Empty pipe OFF Fixed* Stop
Err-21 MULTIRANGE ERROR Multi range err OFF Fixed* Stop
Err-22 COIL SHORT Excitation failure OFF Fixed* Stop
Err-23 EEPROM NO SET EEPROM no set OFF 21.6mA or more** Stop
Err-24 EEPROM DEFAULT EEPROM default OFF Fixed* Stop
Err-25 Dev id not entered OFF Fixed* Stop
Display detail
BRAIN Terminal
HART
Communicator
Indicator
of SE
Alarm
Contact
Output
Current Output
Total/Pulse
Output
Note) *Depending on the setting in “4-20 ALM OUT (Analog alarm out)”.
**When optional code /C1 is selected, current output becomes 2.4mA or less.