IM 01E20D01-01E
5-2
5. MAINTENANCE
5.2 Removing, Cleaning,
and Installing Re-
placeable Electrodes
(General-Purpose
Use Type Only)
5.2.1 Removing Replaceable Elec-
trodes
(1) The following tools are required to replace the
electrodes:
Special tool for removing and installing elec-
trodes (F9807SK): Optional
Torque wrench or torque driver (nominal size 12)
Phillips screwdriver
F0503.EPS
F9807SK
(2) Turn off the power to the flowmeter.
(3) Drain the fluid from the interior of the flowtube.
WARNING
• Never fail to drain the fluid from the interior of
the flowtube.
• If the fluid is at a high temperature, the instru-
ment itself may also be hot. Be careful not to
get burned when removing the electrodes.
• When the process fluid is hazardous to hu-
mans, exercise caution to avoid coming into
contact with it, and avoid inhaling any residual
gas.
(4) Using end A of the tool, open the electrode cover.
F0504.EPS
End B
End A
(5) Hold down the electrode lead wire and terminal
lug and remove the mounting screw using the
Phillips screwdriver.
F0505.EPS
Electrode lead wire:
Red: Electrode A
White: Electrode B
NOTE
Always hold down the lead wire and terminal lug
when removing the mounting screw.