<9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>
9-2
IM 01E30D01-01EN
CongurationTool,downloadthemfromtheofcial
HART programming sites, otherwise, contact the
respectivevendorsoftheCongurationToolforits
upgrade information.
1. ConrmationofthedevicerevisionforAXR
ConnecttheCongurationTooltotheAXR.
The device revision can be checked as follows.
[Root Menu]→Review→Review4→Flddevrev
2. Conrmationofthedevicerevisionforthe
HARTCongurationTool
(1) TurnonthepoweroftheCongurationTool
under the standalone condition.
(2) ConrmtheinstalledDDrevisionin
accordance with the procedure of the
CongurationTool.Refertoitsmanualhow
toconrmitindetail.
9.3 InterconnectionbetweenAXRandHARTCongurationTool
TheHARTCongurationToolcaninterfacewiththeAXRfromthecontrolroom,theAXRsite,oranyother
wiringterminationpointintheloop,providedthereisaminimumloadresistanceof250Ωbetweenthe
connectionandthereceivinginstrument.Tocommunicate,itmustbeconnectedinparallelwiththeAXR,and
the connections must be non-polarized. Figure 9.3.1 illustrates the wiring connections for a direct interface at
theAXRsite.TheHARTCongurationToolcanbeusedforremoteaccessfromanyterminalstripaswell.
AXR
F0903.ai
+
SUPPLY
DO
USB
FieldMate
modem
Relaying
terminals
Distributor
Control room
Terminal
board
USB
PC/FieldMate
Figure 9.3.1 Connecting the HART Communicator
WARNING
Be sure to set parameters as “Protect” on the
writeprotectfunctionafternishofparameter
setting work.
Refer to “Chapter 6 Menu P: Parameter
Protection items” and section “11.3.2” how to use
the write protect function in detail.
IMPORTANT
Ifthepowerofowmeteristurnedoffwithin30
seconds after parameters have been set, these
settings will be canceled. Accordingly, please
keep the power on for at least 30 seconds after
setting parameters.
9.2.2 Device Type Manager (DTM) and
Device Revision
WhenconguretheparametersusingFieldMate,
use the DTM (Device Type Manager) referring to
the following table.
MODEL DTM
Name
Device
Type
Device
Revision
AXR
A
XRHART7
DTM
AXR
(0x3757)
10
The device revision can be checked as follows.
[Root Menu]→Review→Review4→Flddevrev
Zobrazit stránku 120