<6. PARAMETER DESCRIPTION>
6-20
IM 01E30D01-01EN
(2) Menu C: Basic Setting items
Menu C principally contains the basic setting items
f
ortheowtube.
NOTE
Inordertoensurethatcorrectowratedatacan
be acquired, it is crucial that the nominal size,
owratespan,andmeterfactorofthecombined
remoteowtubeareset.Incaseswherethe
AXRintegralowmeterisordered,thenominal
size and meter factor will be set upon shipment
from the manufacturing plant, and these will not
require additional setting.
Ifaowratespanwasspecieduponordering,
this will be set before shipment. If this is not
the case, however, it will be necessary for the
appropriate value to be set by the user.
[C10: Tag No] Setting of the tag number
The setting for this parameter corresponds to
one of the ordered items.
Up to a maximum of 16 characters can be entered
for the display unit. In case of using the HART
c
ongurationTool,uptoamaximumof8characters
can be entered. For more details regarding the
actual characters that can be used, refer to Section
5.2.2: Setting Mode.
NOTE
Ifatagnumberisspecieduponordering,this
parameter is set up before shipment; however,
if this is not the case, it will be necessary for the
appropriate value to be set up by the user.
[C11: Flow Damping] Setting of the damping time
Refer to the description of parameter B20.
[C20: Low MF (IEL)] Setting of the low-frequency
meter factor (Excitation current (Low))
This parameter sets the low-frequency meter factor
for excitation current (Low).
[C21: High MF (IEL)] Setting of the high-frequency
meter factor (Excitation current (Low))
This parameter sets the high-frequency meter
factor for excitation current (Low).
[C22: Low MF (IEM)] Setting of the low-frequency
meter factor (Excitation current (Middle))
This parameter sets the low-frequency meter factor
as required for excitation current (Middle).
[C23: High MF (IEM)] Setting of the high-frequency
meter factor (Excitation current (Middle))
This parameter sets the high-frequency meter
factor for excitation current (Middle).
[C24: Low MF(IEH)] Setting of the low-frequency
meter factor (Excitation current (High))
This parameter sets the low-frequency meter factor
for excitation current (High).
[C25: High MF(IEH)] Setting of the high-frequency
meter factor (Excitation current High))
This parameter sets the high-frequency meter
factor for excitation current (High).
NOTE
Meter Factor Settings
Meter factors
F0601.ai
42
(1) The values of the meter factors which are
markedatMETERFACTOReldsonthe
dataplateoftheAXRaresetintheAXRat
the manufacturing plant.
(2) The meter factors are crucial in ensuring
that the electromotive force is correctly
inproportiontotheowvelocityandare
determined at the manufacturing plant by
actual-owcalibration.
[C30: Nominal Size Unit] Setting of the nominal
size units
This parameter selects the units used for setting of
the nominal size.
[C31: Nominal Size] Setting of the nominal size
T
hisparametersetsthenominalsizeofowtube.
[C40: Base Flow Unit]Selectionofowunitsfor
theowratespan
Refer to the description of parameter B21.