<3. INSTALLATION>
3-10
IM 01E30D01-01EN
Table3.3.8 CenteringDeviceIdentication(AXRLaylengthcode2,PFAlining)
Flange
ratings
Size
mm (inch)
JIS ANSI DIN
10K 20K
F12
(75M)
150 300 PN10 PN16 PN40
50 (2.0) B B — B F — — F
65 (2.5) B B — B G — F —
80 (3.0) B F H F C — G —
100 (4.0) B F H C H — F —
150 (6.0) C D D C E — C —
200 (8.0) C D D D E C C —
*
:Eachcenteringdeviceisengravedwithacharacterasidentication.
Table 3.3.9 Inner Diameters of Grounding Ring,
Outer Diameter for Effective Sealing
Be sure to choose a gasket with an inner and outer
diameter that does not protrude inside the piping.
If the inner diameter of the gasket is too large,
orouterdiameterofthegasketistoosmall,uid
leakage may result.
Lay length code1 (Flange, Wafer type) Unit : mm (in.)
Size
Inner diameter of
ground ring[øA]
Outer diameter for
effective sealing[øB]
25 (1.0) 28 (1.10) 53 (2.09)
40 (1.5) 41 (1.61) 71 (2.80)
50 (2.0) 53 (2.09) 84 (3.31)
65 (2.5) 66 (2.60) 103 (4.06)
80 (3.0) 77 (3 .03) 114 (4.49)
100 (4.0) 102 (4.02) 140 (5.51)
150 (6.0) 146.1 (5.75) 190 (7.48)
200 (8.0) 193.6 (7.62) 240 (9.45)
L
ay length code2 (Wafer type) Unit : mm (in.)
Size
Inner diameter of
ground ring[øA]
Outer diameter for
effective sealing[øB]
25 (1.0) 28 (1.10) 53 (2.09)
40 (1.5) 41 (1.61) 71 (2.80)
50 (2.0) 53 (2.09) 84 (3.31)
65 (2.5) 66 (2.60) 103 (4.06)
80 (3.0) 77 (3 .03) 114 (4.49)
100 (4.0) 102 (4.02) 140 (5.51)
150 (6.0) 140.7 (5.54) 190 (7.48)
200 (8.0) 188.9 (7.44) 240 (9.45)
Size of Inner Diameter of Grounding ring and
Outer Diameter for Effective Sealing :
lining
øB øA
AXR Flowtube
Grounding ring
Gasket for
User’s flange
3.3.3 Nominal Diameter 25 mm (1.0 in.) to
200 mm (8.0 in.), Flange Type
Useboltsandnutsincompliancewiththeange
ratings. Be sure to choose a gasket with inner
and outer diameters that does not protrude
inside the piping (refer to Table 3.3.9). If the
inner diameter of the gasket is too large, or outer
diameterofthegasketistoosmall,uidleakage
may result.
(1) Mounting Direction
Mounttheowmetersothattheowdirectionofthe
uidtobemeasuredisinlinewiththedirectionof
thearrowmarkontheowmeter.
If it is impossible to match the direction of
the arrow mark, the direction of the electrical
connection can be changed. Refer to Section
10.1 to do this properly.
Incasetheuidbeingmeasuredowsagainst
the arrow direction, refer to the parameter J20:
Flow Direction in this user’s manual.
(2) Tightening Nuts
Tighten the bolts according to the torque values for
the metal piping in Table 3.3.10. For PVC piping,
select an optional code of GA, GC, or GD, use
rubber gaskets and tighten the nuts to the torque
values for the PVC piping in Table 3.3.11.
F
orpermeableuids(suchasnitricacid,
hydrouoricacid,orsodiumhydrateathigh
temperatures), tighten the nuts according to the
torque values in Table 3.3.12.